Randomized, Crossover Trial of Control-IQ Technology with a Lower Treatment Range and a Modified Meal Bolus Module in Adults, Adolescents, Children, and Preschoolers with Varying Levels of Baseline Glycemic Control

Diabetes Technol TherPublished Randomised crossover trial

Authors' conclusion

The lower treatment range and late bolus feature of the modified Control-IQ system were safe for use in all age-groups. TIR and TITR improved with LR regardless of baseline HbA1c.

Study context

Population
Adults, adolescents, children and preschoolers with T1D, aged 3–57 yearscompleted the study
Participants
72 adults
Comparison
Modified Control-IQ with lower treatment range (LR, 90-130 mg/dL) vs standard range (SR, 112.5-160 mg/dL) vs run-in (Control-IQ v1.5)
Methodological detailsmulticenter, at home in random order, Two late bolus meal challenges per 2-week period, bolusing 45 min after meals with and without a new late bolus feature
  • Multicenter
  • At home in random order
  • Two late bolus meal challenges per 2-week period, bolusing 45 min after meals with and without a new late bolus feature

Primary result

As reported in the abstract

TIR 70-180 mg/dL

Difference

+3.1%

P < 0.001

LR vs Run-in

Difference

+2.1%

P < 0.001

LR vs SR

LR

68.0%

Further findings

OutcomeDifferenceP value
Time in tight range (TITR) 70-140 mg/dLLR vs Run-in+3.3%P < 0.001
Time in tight range (TITR) 70-140 mg/dLLR vs SR+4.0%P < 0.001

Time in tight range (TITR) 70-140 mg/dL

Similar improvements with LR for time >180 mg/dL and mean glucose

Time in range (TIR) and TITR by baseline hemoglobin A1c (HbA1c)

Lower baseline HbA1c: highest TIR and TITR with LR; higher baseline HbA1c: greatest improvements in TIR and TITR

Safety

There were no diabetic ketoacidosis or severe hypoglycemia events
All meal challenges were completed safely

Original publication

Randomized, Crossover Trial of Control-IQ Technology with a Lower Treatment Range and a Modified Meal Bolus Module in Adults, Adolescents, Children, and Preschoolers with Varying Levels of Baseline Glycemic Control. Diabetes Technol Ther. Published 26 November 2024. doi:10.1089/dia.2024.0501

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.