Diabetes technology

Control-IQ Technology Use in Individuals With High Insulin Requirements: Results From the Multicenter Higher-IQ Trial

J Diabetes Sci TechnolPublished Single-arm prospective study

Authors' conclusion

Control-IQ technology version 1.5, with wider range of weight and TDI input and enhancements for users with high insulin requirements, was safe in individuals with T1D in this study.

Study context

Population
Adults with T1D using continuous subcutaneous insulin infusion (CSII) with at least one basal rate >3 units/h; mean age 39.9 years, 41.2% female, diabetes duration 21.8 yearsall completed
Participants
34 adults
Outcome
Safety events:
  • Severe hypoglycemia
  • DKA
Follow-up
13 weeks
Methodological detailsmulticenter
  • Multicenter

Main finding

As reported in the abstract

  • There were no severe hypoglycemia or diabetic ketoacidosis (DKA) events

Observational data: associations, not effects.

Further findings

Time in range 70-180 mg/dL · over the 13 weeks

Time in range 70-180 mg/dL
64.8%

Time <70 mg/dL · over the 13 weeks

Time <70 mg/dL
1.0%

Hemoglobin A1c (HbA1c) (central laboratory) · baseline to 13 weeks

HbA1c at baseline
7.69%
HbA1c at 13 weeks
6.87%
HbA1c change
−0.82%, P < .001

Safety

Severe hypoglycemia and DKA · 13 weeks

There were no severe hypoglycemia or DKA events

Original publication

Control-IQ Technology Use in Individuals With High Insulin Requirements: Results From the Multicenter Higher-IQ Trial. J Diabetes Sci Technol. Published 5 March 2024. doi:10.1177/19322968241234072

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.