Osteoporosis, bone & mineral

The quantification of bone mineral density using photon counting computed tomography and its implications for detecting bone remodeling

J Bone Miner ResPublished Methodological imaging study

Main finding · Tt.vBMD accuracy, PCCT vs HR-pQCTResult — go to the resultbias 7.46 mgHA/cm³

Authors' conclusion

Based on the cadaver data, we conclude that PCCT can be used to quantify vBMD and bone turnover.

Study context

Population
Eight cadaveric forearms (PCCT twice, HR-pQCT once) and dominant forearm of two volunteers (PCCT twice); carpals delineated8 cadaveric forearms · 2 volunteers
Comparison
PCCT vs HR-pQCT
Outcome
Tt.vBMD accuracy, PCCT vs HR-pQCT
Methodological detailsex vivo cadaveric forearms, in vivo volunteers, phantom calibration
  • Ex vivo cadaveric forearms
  • In vivo volunteers
  • Phantom calibration

Main finding

As reported in the abstract

bias

7.46 mgHA/cm³

Pearson correlation of paired Tt.Volumetric BMD (vBMD)
0.986

Further findings

Voxel-level repeatability of repeated PCCT scans (ex vivo)

mgHA/cm³

Bias
17.66
SE
96.23

Least detectable remodeling

Least detectable remodeling threshold
250 mgHA/cm³
Voxels incorrectly classified as newly added or resorbed bone (ex vivo)
0.37%
Voxels incorrectly classified (in vivo)
0.97%

Original publication

The quantification of bone mineral density using photon counting computed tomography and its implications for detecting bone remodeling. J Bone Miner Res. Published 4 October 2024. doi:10.1093/jbmr/zjae163

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.