Diabetes technology

Reliability of Handheld Blood Glucose Monitors in Neonates: Trustworthy Arterial Readings but Capillary Results Warrant Caution for Hypoglycemia

J Diabetes Sci TechnolPublished Method-comparison study

Authors' conclusion

Blood glucose monitor testing provided reliable results from arterial blood. However, users should interpret hypoglycemic results obtained from capillary blood with caution.

Study context

Population
Preterm and ill neonates in neonatal intensive care units (NICUs)
Participants
40 participants
Comparison
Capillary and arterial blood glucose monitors (BGM) vs arterial plasma YSI 2300 Stat Plus glucose
Outcome
Differences in glucose concentrations between capillary BGM and arterial YSI, and between arterial BGM and YSI aliquots of the same sample

Main finding

As reported in the abstract

Results within CLSI and FDA tolerance criteria

Capillary BGM results outside tolerance criteria (average)

27.5%

Original publication

Reliability of Handheld Blood Glucose Monitors in Neonates: Trustworthy Arterial Readings but Capillary Results Warrant Caution for Hypoglycemia. J Diabetes Sci Technol. Published 20 October 2023. doi:10.1177/19322968231207861