Diabetes technology

Clinical Use of Continuous Glucose Monitoring in Critically Ill Pediatric Patients with Diabetic Ketoacidosis

Diabetes Technol TherPublished Diagnostic accuracy study

Authors' conclusion

The severity of DKA has a major effect on reducing the accuracy of the CGM, especially during the first several days in the intensive care unit. The reduced accuracy appears to be related to acidosis, as reflected in the serum bicarbonate levels.

Study context

Population
Pediatric patients with diabetic ketoacidosis (DKA) in the pediatric intensive care units (PICU); mean age 10.98 ± 4.20 years
Participants
399 matched pairs
Comparison
Continuous glucose monitoring (CGM) vs point-of-care capillary glucose
  • Medtronic Guardian Sensor 3n = 331
  • Dexcom G6n = 41
  • Abbott FreeStyle Libre 1n = 27
Outcome
Accuracy of CGM vs POC (MARD)

Performance

As reported in the abstract

MARD
OutcomeMedtronic Guardian Sensor 3Dexcom G6Abbott FreeStyle Libre 1
MARD13.40%11.12%11.33%
Overall mean absolute relative difference (MARD)
13.02%
Surveillance error grid (SEG) zones A and B
98.5%
Bland-Altman mean difference
15.5 mg/dL
Pearson's correlation coefficient [r2]
0.76, P < 0.0001

Further findings

MARD by sensor change · MARD significantly lower without sensor change

P = 0.048

  • MARD by sensor change: 3 patients in sensor change group

Correlation of serum bicarbonate with POC-CGM values

Significant negative correlation (r)
−0.34, P < 0.001

Original publication

Clinical Use of Continuous Glucose Monitoring in Critically Ill Pediatric Patients with Diabetic Ketoacidosis. Diabetes Technol Ther. Published 17 July 2023. doi:10.1089/dia.2023.0012

Composed by AI from the published abstract and checked against it. How Briefs are made →