Accuracy of Factory-Calibrated Continuous Glucose Monitors in Critically Ill Patients Receiving Intravenous Insulin: A Prospective Clinical Trial of Two Leading Systems

J Diabetes Sci TechnolPublished Prospective clinical trial

Authors' conclusion

Factory-calibrated continuous glucose monitors may require accuracy validation and per-patient calibration for inpatient use in critically ill patients.

Study context

Population
Critically ill hospitalized patients requiring continuous intravenous insulin infusion
Participants
40 participants
Exposure
Dexcom G6 Pro (G6P) vs FreeStyle Libre Pro (FLP) against reference serum (Lab) and point-of-care (POC) glucose
Outcome
CGM accuracy (MARD) vs reference Lab and POC glucose
Methodological detailsnonrandomized, open-label, clinically blinded
  • Nonrandomized
  • Open-label
  • Clinically blinded

Main finding

As reported in the abstract

MARD vs reference glucose
GroupValueMean absolute relative difference (MARD) vs Lab (1015 pairs)MARD vs POC (2369 pairs)MARD vs Lab (1006 pairs)MARD vs POC (2353 pairs)
G6PMARD 22.7%22.7%22.9%
FLPMARD 25.2%25.2%27.0%

Value

G6PMARD 22.7%
FLPMARD 25.2%

Mean absolute relative difference (MARD) vs Lab (1015 pairs)

G6P22.7%
FLP— not stated in the abstract

MARD vs POC (2369 pairs)

G6P22.9%
FLP— not stated in the abstract

MARD vs Lab (1006 pairs)

G6P— not stated in the abstract
FLP25.2%

MARD vs POC (2353 pairs)

G6P— not stated in the abstract
FLP27.0%

Observational data: associations, not effects.

Original publication

Accuracy of Factory-Calibrated Continuous Glucose Monitors in Critically Ill Patients Receiving Intravenous Insulin: A Prospective Clinical Trial of Two Leading Systems. J Diabetes Sci Technol. Published 8 May 2025. doi:10.1177/19322968251338865

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.