Diabetes technology

Accuracy of Dexcom One+ in Patients with Diabetes or Stress Hyperglycemia Hospitalized in Cardiac Intensive Care Unit

Diabetes Technol TherPublished Diagnostic accuracy study

Authors' conclusion

In patients in CICU, the Dexcom One+ system showed acceptable accuracy and could support glucose monitoring.

Study context

Population
Patients with known diabetes or hyperglycemia on admission >140 mg/dL hospitalized in the cardiac ICU for acute coronary syndrome and/or HF, September 2024 to May 2025; 39 with previously diagnosed T2D, aged 73.5 ± 9.6 years, mean HbA1c 7.1 ± 1.3%999 paired samples
Participants
48 patients
Comparison
Dexcom One+ continuous glucose monitoring (CGM) vs standard blood glucose measurements
Outcome
Accuracy by mean absolute relative difference (MARD)

Performance

As reported in the abstract

Accuracy: MARD vs reference blood glucose

Overall mean absolute relative difference (MARD)

11.6%

95% CI 10.9–12.2

Further findings

Clinical performance: Consensus Error Grid analysis

Readings in Zone A
93.5%
Readings in Zone B
6.1%
Readings in Zone C
0.4%
No readings were found in Zone D + E

Safety

Skin reactions

Mild hematoma at insertion site
1

Original publication

Accuracy of Dexcom One+ in Patients with Diabetes or Stress Hyperglycemia Hospitalized in Cardiac Intensive Care Unit. Diabetes Technol Ther. Published 23 January 2026. doi:10.1177/15209156251407717

Composed by AI from the published abstract and checked against it. How Briefs are made →