Diabetes technology

Real-World Evidence of Omnipod ® 5 Automated Insulin Delivery System Use in 69,902 People with Type 1 Diabetes

Diabetes Technol TherPublished Retrospective cohort study

Authors' conclusion

These glycemic outcomes from a large and diverse sample of nearly 70,000 children and adults demonstrate effective use of the Omnipod 5 System under real-world conditions.

Study context

Population
US Omnipod 5 System users aged ≥2 years with T1D and sufficient data (≥90 days of data; ≥75% of days with ≥220 CGM readings/day), Insulet device and person-reported datasets as of July 2023
Participants
69,902 participants
Exposure
Average glucose targets of 110, 120 and 130-150 mg/dL
  • 110 mg/dLn = 37,640
Outcome
Median time in range (TIR) (70-180 mg/dL)
Methodological detailsreal-world, subgroup analyses, stratification by baseline characteristics
  • Real-world
  • Subgroup analyses
  • Stratification by baseline characteristics

Main finding

As reported in the abstract

Observational data: associations, not effects.

Further findings

Median

OutcomeChildren and adolescents (2-17 years)Adults (≥18 years)
Median TIR, average glucose target 110 mg/dL, by age65.0%69.9%

Subgroups transitioning from Omnipod DASH or MDI, and Medicaid/Medicare users

Favorable glycemic outcomes among these groups

Original publication

Real-World Evidence of Omnipod ® 5 Automated Insulin Delivery System Use in 69,902 People with Type 1 Diabetes. Diabetes Technol Ther. Published 16 February 2024. doi:10.1089/dia.2023.0578

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.