Diabetes technology

Investigation of the Effect of 70 Potential Interferents on Measurement Results of Two Blood Glucose Monitoring Systems

J Diabetes Sci TechnolPublished Interference screening study

Authors' conclusion

The BGMS investigated in this study were found to be robust with respect to the 70 evaluated potential interferents. Interference effects were observed only for N-Acetyl-L-cysteine. Extensive evaluations of potential interfering factors can make an important contribution to ensure reliability of BGMS.

Study context

Population
Venous whole blood samples; 70 potential interferents expected to be common in people with diabetes70 potential interferents
Comparison
Test samples with potential interferents vs control samples, for two blood glucose (BG) monitoring systems
Outcome
Interference effects as normalized biases between test and control samples; acceptance within ±10 mg/dL (glucose ≤100 mg/dL) or ±10% (>100 mg/dL)
Methodological detailspaired-difference approach based on ISO 15197:2013 and CLSI guideline EP07, three reagent system lots per system, venous whole blood samples at two glucose concentrations
  • Paired-difference approach based on ISO 15197:2013 and CLSI guideline EP07
  • Three reagent system lots per system
  • Venous whole blood samples at two glucose concentrations

Recommendations

As reported in the abstract

Normalized bias within predefined acceptance criteria

Potential interferents for which both systems met acceptance criteria, for each of the three reagent system lots

69 of 70

Original publication

Investigation of the Effect of 70 Potential Interferents on Measurement Results of Two Blood Glucose Monitoring Systems. J Diabetes Sci Technol. Published 15 February 2024. doi:10.1177/19322968241231294

Composed by AI from the published abstract and checked against it. How Briefs are made →