Diabetes technology

Accuracy and Safety of the 15-Day CareSens Air Continuous Glucose Monitoring System

Diabetes Technol TherPublished Diagnostic accuracy study

Authors' conclusion

The CareSens Air demonstrated accurate performance across the wide glycemic range and was well tolerated during the 15-day sensor use period.

Study context

Population
Adults with T1D or T2D requiring intensive insulin therapy, 4 sites in South Korea
Participants
83 adults10,029 CGM-YSI pairs
Comparison
CareSens Air continuous glucose monitoring (CGM) vs YSI reference
  • CareSens Airn = 83
Outcome
Accuracy: proportion of CGM values within 15% of YSI values ≥100 mg/dL or within 15 mg/dL of YSI values <100 mg/dL (%15/15), with %20/20, %30/30, %40/40 agreement rates; MARD vs YSI
Follow-up
15 days
Methodological detailsmulticentre, single-group, Four 8-h clinic sessions on Day 1, 5 ± 1, 10 ± 1, and 15
  • Multicentre
  • Single-group
  • Four 8-h clinic sessions on Day 1, 5 ± 1, 10 ± 1, and 15

Performance

As reported in the abstract

Overall accuracy vs YSI (10,029 CGM-YSI matched pairs)
Overall mean absolute relative difference (MARD)
10.42%
Overall %15/15 accuracy
78.55%
Overall %20/20 accuracy
89.04%
Overall %30/30 accuracy
96.47%
Overall %40/40 accuracy
98.87%

Further findings

Consensus error grid analysis

CGM values in Zone A or B
99.92%
Zone A
89.83%
Zone B
10.09%

%20/20 accuracy by day

Day 1
88.11%
Day 3-5
90.11%
Day 8-10
92.09%
Day 15
85.63%

Safety

Serious adverse events
0

Original publication

Accuracy and Safety of the 15-Day CareSens Air Continuous Glucose Monitoring System. Diabetes Technol Ther. Published 13 February 2024. doi:10.1089/dia.2023.0468