Performance of Subcutaneous Continuous Glucose Monitoring in Adult Critically Ill Patients Receiving Vasopressor Therapy

Diabetes Technol TherPublished Diagnostic accuracy study

Authors' conclusion

In our ICU cohort of patients receiving vasopressor infusion, subcutaneous CGM demonstrated acceptable overall numerical and clinical accuracy. However, suboptimal accuracy may occur outside glucose ranges of 6-10 mmol/L (108-180 mg/dL), during higher dose vasopressor infusion, and during the first 36 h after CGM start.

Study context

Population
Adult ICU patients receiving mechanical ventilation, IV insulin infusion and vasopressor therapy; 18 with T2D; median (IQR) maximum noradrenaline equivalent infusion rate 0.18 (0.08–0.33) µg/kg/min2,946 paired samples
Participants
40 patients
Comparison
Continuous glucose monitoring (CGM) vs arterial blood glucose
Outcome
Numerical accuracy by mean absolute relative difference (MARD) vs ABG; MARD <14% acceptable

Performance

As reported in the abstract

Accuracy vs ABG
Test or modelValueMean absolute relative difference (MARD)CGM readings within Clarke Error Grid (CEG) zones A and B
CGM12.7% (95% CI 10.7–15.3)12.7%99.8%

CGM

Value12.7% (95% CI 10.7–15.3)
Mean absolute relative difference (MARD)12.7%
CGM readings within Clarke Error Grid (CEG) zones A and B99.8%

Further findings

MARD by arterial blood glucose (ABG) strata and noradrenaline dose

MARD ≥14% when ABG outside target range (6-10 mmol/L [108-180 mg/dL]) and with noradrenaline equivalent infusion rates above 0.10 µg/kg/min

MARD over time since CGM start

Accuracy improved with time, MARD <14% after 36 h

Feasibility: interrupted sensor readings due to signal loss

Episodes of interrupted sensor readings
4
Duration of interruptions
5 to 20 min

Safety

Skin reactions related to sensor insertion or sensor adhesives (tolerability)

No skin reaction related to sensor insertion or sensor adhesives

Original publication

Performance of Subcutaneous Continuous Glucose Monitoring in Adult Critically Ill Patients Receiving Vasopressor Therapy. Diabetes Technol Ther. Published 31 May 2024. doi:10.1089/dia.2024.0035

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.