Diabetes technology

Determinants of Continuous Glucose Monitoring Accuracy in Critically Ill Patients Requiring Mechanical Ventilation and Vasopressor Support

Diabetes Technol TherPublished Secondary analysis of a prospective observational study

Authors' conclusion

In critically ill patients receiving mechanical ventilation and vasopressor therapy, CGM accuracy improves substantially with increasing sensor wear time and is independently associated with arterial pH.

Study context

Population
Critically ill adults receiving IV insulin, mechanical ventilation and vasopressor therapy; Continuous glucose monitoring (CGM) with Dexcom G6®2,946 paired arterial blood glucose (ABG)–CGM values
Participants
40 adults
Primary endpoint
Percentage absolute relative difference (ARD) between CGM and arterial blood glucose
Methodological detailsmultivariable generalized linear mixed models with patient-level random intercepts, B-splines for continuous covariates
  • Multivariable generalized linear mixed models with patient-level random intercepts
  • B-splines for continuous covariates

Primary result

As reported in the abstract

ARD
Median absolute relative difference (ARD) (IQR 5.2-18.8)
10.8%
Mean ARD
12.7 % (95% CI 10.7-15.3)
Sensor wear time independently associated with ARD and both categorical accuracy metrics
P < 0.001
pH independently associated with ARD and both categorical accuracy metrics
P ≤ 0.02
Each 0.1-unit increase in pH: higher ARD (linear-term sensitivity analysis; P = 0.009)
1.5 percentage-point

Secondary outcomes

Readings meeting ISO 15197:2013 and CLSI POCT12-A3 criteria

Readings meeting ISO criteria
64.5%
Readings meeting CLSI criteria
56.0%
Each 0.1-unit increase in pH: lower odds of meeting CLSI criteria (linear-term sensitivity analysis; P = 0.001)
32%
Noradrenaline-equivalent (NE)-dose independently associated with ISO-defined accuracy
P = 0.006
ABG independently associated with ISO-defined accuracy
P < 0.001
Probability of meeting ISO/CLSI criteria increased during first 50 h after insertion, then stabilized
Spline analyses suggested reduced accuracy at higher NE-doses (>0.6 µg/kg/min) and lower ABG (<5 mmol/L [<90 mg/dL]); limited observations in these ranges

Original publication

Determinants of Continuous Glucose Monitoring Accuracy in Critically Ill Patients Requiring Mechanical Ventilation and Vasopressor Support. Diabetes Technol Ther. Published 20 August 2026. doi:10.1177/15209156261479816

Composed by AI from the published abstract and checked against it. How Briefs are made →