Diabetes technology

Accuracy of an Interstitial Continuous Glucose Management System in Critically Ill Patients: A Prospective Observational Study

J Diabetes Sci TechnolPublished Prospective observational study

Authors' conclusion

Dexcom G7 showed acceptable accuracy during stable ICU conditions, but performance deteriorated during rapid glucose excursions.

Study context

Population
Critically ill adults with stress hyperglycemia, no prior diabetes, expected intensive care unit (ICU) stay ≥5 days
Participants
20 adults
Exposure
Dexcom G7 continuous glucose monitoring (CGM) vs arterial blood glucose
Outcome
CGM accuracy vs arterial reference glucose, outside IVGTT
Follow-up
First 10 ICU days
Methodological detailsancillary study of the Determination of Insulin Resistance in ICU (DIRICU) study, post hoc 20-minute temporal shift analysis
  • Ancillary study of the DIRICU study
  • Post hoc 20-minute temporal shift analysis

Main finding

As reported in the abstract

Paired values
716
R2
.58
Mean bias
+11 mg/dL
Mean absolute relative difference (MARD)
17%
Values in Clarke zones A/B
98.9%

Observational data: associations, not effects.

Further findings

CGM accuracy vs arterial reference glucose, during intravenous glucose tolerance tests (IVGTT)

Paired values
1018
R2
.028
Mean bias
−34 mg/dL
MARD
24%
Values in zones A/B
80.3%
Values in zone D
18.6%
  • Accuracy deteriorated markedly

CGM accuracy during IVGTT after 20-minute temporal shift

R2
.34
MARD
19%
Values in zones A/B
94.7%
  • Post hoc analysis

Safety

Sensor-related adverse events
0

Original publication

Accuracy of an Interstitial Continuous Glucose Management System in Critically Ill Patients: A Prospective Observational Study. J Diabetes Sci Technol. Published 21 September 2026. doi:10.1177/19322968261486451

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.