Diabetes technology

Accuracy and Feasibility of Real-time Continuous Glucose Monitoring in Critically Ill Patients After Abdominal Surgery and Solid Organ Transplantation

Diabetes CarePublished Accuracy and feasibility study

Authors' conclusion

Our study shows clinically applicable accuracy and reliability of Dexcom G6 CGM in postoperative ICU patients and a feasible alternative sensor placement site.

Study context

Population
ICU patients after pancreatic surgery or solid organ transplantation (liver, pancreas, islets of Langerhans, kidney)1,546 paired values
Participants
61 patients
Comparison
Real-time continuous glucose monitoring (CGM) vs arterial blood glucose as reference
Outcome
Feasibility and accuracy of real-time CGM (MARD, bias, modified Bland-Altman plot, surveillance error grid) vs ABL

Performance

As reported in the abstract

Sensor accuracy of real-time CGM vs ABL
Mean absolute relative difference (MARD)
9.4%
Relative bias
1.4%
Surveillance error grid zone A
92.8%
Surveillance error grid zone B
6.1%
Surveillance error grid zone C
1.2%
Paired glucose values from CGM and ABL
1,546

Further findings

Feasibility: active sensor use

Active sensor use
95.1%

Glycaemic control

Median TIR
78%
Time spent in hypoglycaemia
<1%

StatStrip glucose meter accuracy vs ABL

StatStrip MARD
5.8%

Original publication

Accuracy and Feasibility of Real-time Continuous Glucose Monitoring in Critically Ill Patients After Abdominal Surgery and Solid Organ Transplantation. Diabetes Care. Published 27 February 2024. doi:10.2337/dc23-1663

Composed by AI from the published abstract and checked against it. How Briefs are made →