Diabetes technology

Performance of continuous glucose monitoring systems ( continuous glucose monitoring (CGMs) ) during commercial flights in T1DM: A within‐subject comparative pilot study

Diabetes Obes MetabPublished Within-subject comparative pilot study

Authors' conclusion

G7 and Libre3 showed consistently higher accuracy and clinical reliability across various phases of flight and ground conditions. Simplera and i3 exhibited greater variability in accuracy. These findings suggest that while modern CGMs could support diabetes management in aviation, individual system performance may differ under specific operational conditions.

Study context

Population
Adults with T1D on commercial flights; age 37.9 ± 12.8 years, HbA1c 52.3 ± 4.9 mmol/mol, diabetes duration 19.25 ± 10.6 years
Participants
20 adults3,473 matched pairs
Comparison
Four CGMs during in-flight and ground conditions
Outcome
Overall mean absolute relative difference (MARD)
Methodological detailsfour sensors worn in parallel
  • Four sensors worn in parallel

Main finding

As reported in the abstract

Overall MARD G7 9.5% ± 3.3% · Libre3 9.6% ± 3.8% · Simplera 19.9% ± 6.2% · i3 15.7% ± 6.6%

%20/20

%10/10

Consensus error grid Zones A and B

Further findings

Prandial state and flight phase had limited influence on the relative performance ranking of the devices

Original publication

Performance of continuous glucose monitoring systems ( CGMs ) during commercial flights in T1DM: A within‐subject comparative pilot study. Diabetes Obes Metab. Published 2 January 2026. doi:10.1111/dom.70418

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.