Performance of three simultaneously used real-time continuous glucose monitoring (rtCGM) systems around a physically active weekend camp in adults with type 1 diabetes: A prospective lab and real‐world study
Authors' conclusion
Dexcom G7 and FreeStyle Libre 3 showed very high accuracy, which could not be found for the Simplera rtCGM system.
Study context
- Population
- Adults with T1D (9 females; age 50 ± 12 years; HbA1c 6.6 ± 0.5%); 3-day lab phase with three 90-min group exercise sessions, then 4-day home phase
- Participants
- 20 adults
- Comparison
- Dexcom G7 vs FreeStyle Libre 3 vs Simplera rtCGM accuracy against capillary blood glucose
- Outcome
- MedARD [IQR] of sensor glucose vs reference capillary blood glucose
Methodological detailslab and real-world phases, simultaneous wear of three rtCGM systems, Duplicate capillary glucose (Contour Next One) every 2 h, 8:00 AM–10:00 PM
- Lab and real-world phases
- Simultaneous wear of three rtCGM systems
- Duplicate capillary glucose (Contour Next One) every 2 h, 8:00 AM–10:00 PM
Performance
As reported in the abstract
MedARD vs reference blood glucose
| Test or model | Value | Comparison points |
|---|---|---|
| Dexcom G7 | median 8.4% (IQR 3.6–14.0) | 723 |
| Libre 3 | median 7.5% (IQR 3.1–13.4) | 723 |
| Simplera | median 13.8% (IQR 6.9–22.5) | 719 |
Value
Dexcom G7median 8.4% (IQR 3.6–14.0)
Libre 3median 7.5% (IQR 3.1–13.4)
Simpleramedian 13.8% (IQR 6.9–22.5)
Comparison points
Dexcom G7723
Libre 3723
Simplera719
- Libre 3 vs Dexcom G7: P = 0.35 — no significant difference
- Dexcom G7 vs Simplera: P < 0.0001 — significant difference
- Libre 3 vs Simplera: P < 0.0001 — significant difference