Performance Assessment of Three Continuous Glucose Monitoring Systems in Adults With Type 1 Diabetes
Authors' conclusion
The results of FBS were encouraging enough to conduct further performance and usability evaluations in a low- and middle-income country. The results of FL2 mainly agreed with existing studies, whereas ADX showed substantial deviations from previously reported results.
Study context
- Population
- Adults with T1D30 adults enrolled
- Comparison
- FiberSense vs FreeStyle Libre 2 vs GlucoRx AiDEX
- FiberSense system (FBS)n = 31 sensors
- FL2n = 70 sensors
- AiDEX (ADX)n = 78 sensors
- Outcome
- Accuracy vs comparator: mean absolute relative difference (MARD) and relative bias
- Follow-up
- 28-day study period
Methodological detailsmainly home-based, three in-clinic sessions, capillary blood glucose comparator, Study mainly at home with three in-clinic sessions over 28 days
- Mainly home-based
- Three in-clinic sessions
- Capillary blood glucose comparator
- Study mainly at home with three in-clinic sessions over 28 days
Main finding
As reported in the abstract
| Group | Value | Relative bias | Sensor-to-sensor variability |
|---|---|---|---|
| FBS | 14.7% | −2.1% | low |
| FL2 | 9.2% | −2.5% | moderate |
| ADX | 21.9% | −18.5% | high |
Value
FBS14.7%
FL29.2%
ADX21.9%
Relative bias
FBS−2.1%
FL2−2.5%
ADX−18.5%
Sensor-to-sensor variability
FBSlow
FL2moderate
ADXhigh
Further findings
Sensor survival probability until end of sensor life
Sensor life FBS 28 days · FL2 14 days · ADX 14 days
Original publication
Performance Assessment of Three Continuous Glucose Monitoring Systems in Adults With Type 1 Diabetes. J Diabetes Sci Technol. Published 22 February 2023. doi:10. 1177/ 19322968231159657