Diabetes technology

Multicenter Evaluation Study Comparing a New Factory-Calibrated Real-Time Continuous Glucose Monitoring System to Existing Flash Glucose Monitoring System

J Diabetes Sci TechnolPublished Diagnostic accuracy study

Study context

Population
Participants wearing 4 study sensors with one in-clinic visit for venous blood reference tests
Participants
120 vs 40
Comparison
AiDEX CGM vs Abbott Libre flash glucose monitoring
  • AiDEXn = 120
  • Abbott Libren = 40reference
Outcome
Accurate readings vs venous blood glucose (continuous glucose error grid analysis (CG-EGA))
Methodological detailsmulticenter, prospective, masked
  • Multicenter
  • Prospective
  • Masked

Performance

As reported in the abstract

Further findings

Overall MARD of AiDEX
9.08% · MARD vs venous blood reference · MARD vs finger capillary BG reference 10.1%

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

Multicenter Evaluation Study Comparing a New Factory-Calibrated Real-Time Continuous Glucose Monitoring System to Existing Flash Glucose Monitoring System. J Diabetes Sci Technol. Published 11 August 2021. doi:10.1177/19322968211037991