Continuous Glucose Monitoring and Recreational Scuba Diving in Type 1 Diabetes: Head-to-Head Comparison Between Free Style Libre 3 and Dexcom G7 Performance
Authors' conclusion
The study reveals Free Style Libre 3 superior accuracy compared with Dexcom G7 in a setting of repetitive scuba diving in T1DM, except for hypoglycemic range. Both sensors fail to achieve accuracy for nonadjunctive use.
Study context
- Population
- Patients with T1D performing repetitive recreational scuba diving under a safety protocol, Diabete Sommerso® association event, 2023202 dives
- Participants
- 13 patients
- Comparison
- Dexcom G7 vs FreeStyle Libre 3 during repetitive scuba diving
- Outcome
- Sensor accuracy: MARD, MeARD and surveillance error grid vs capillary glucose
Methodological detailshead-to-head comparison, capillary glucose reference standard
- Head-to-head comparison
- Capillary glucose reference standard
Performance
As reported in the abstract
Sensor accuracy vs capillary glucose
MARD
MeARD
SEG no-risk zone
Further findings
| Outcome | Dexcom G7 | Free Style Libre 3 |
|---|---|---|
| Hypoglycemia identification | ||
| Diagnostic OR | 58.95 | 254.10 |
Original publication
Continuous Glucose Monitoring and Recreational Scuba Diving in Type 1 Diabetes: Head-to-Head Comparison Between Free Style Libre 3 and Dexcom G7 Performance. Diabetes Technol Ther. Published 11 June 2024. doi:10. 1089/ dia. 2024. 0126