Diabetes technology

Continuous Glucose Monitoring for Patients with COVID-19 Pneumonia: Initial Experience at a Tertiary Care Center

Endocr PractPublished Observational study

Authors' conclusion

With twice daily, on-demand calibration, the inpatient CGM use was safe for insulin dosing, decreasing the frequency of glucometer fingersticks. For glucose levels >70 mg/dL, CGMs showed adequate accuracy, without interference from vital and laboratory values.

Study context

Population
Adults hospitalized with COVID-19 and hyperglycemia812 paired values
Participants
20 adults
Exposure
CGM vs glucometer readings
Outcome
CGM accuracy against glucometer, expressed as mean absolute relative difference (MARD)
Methodological detailssingle-centre, tertiary care center, remotely monitored CGM
  • Single-centre
  • Tertiary care center
  • Remotely monitored CGM

Main finding

As reported in the abstract

Aggregate MARD of CGM vs glucometer

Aggregate mean absolute relative difference (MARD)

10.4%

MARD for time in range

11.4%

MARD for grade 1 hyperglycemia

9.4%

MARD for grade 2 hyperglycemia

9.1%

Observational data: associations, not effects.

Further findings

MARD correlation with vital and laboratory values

No MARD correlation with mean arterial blood pressure, oxygen saturation, daily hemoglobin or glomerular filtration rates

Frequency of glucometer measurements · after CGM placement

Glucometer measurements per day decreased from 5 to 3 and then 2

Safety

CGM clinical reliability (Clarke error grid)

CGM values within the "safe" zones of Clarke error grid
99.7%

Original publication

Continuous Glucose Monitoring for Patients with COVID-19 Pneumonia: Initial Experience at a Tertiary Care Center. Endocr Pract. Published 22 December 2022. doi:10.1016/j.eprac.2022.12.012

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.