Diabetes technology

Comparison of continuous glucose monitoring with self-monitoring of blood glucose in type 1 diabetes in the changing atmospheric pressures in aviation: a hypobaric flight simulation

DiabetologiaPublished Mechanistic study

Authors' conclusion

The performance of CGM systems and SMBG are comparable during flight-associated atmospheric pressure changes. All tested measurement devices for CGM and SMBG were suitable for diabetes-care-based decisions during flight simulation.

Study context

Population
Insulin pump users with T1D; 3 continuous glucose monitoring (CGM) brands and 2 self-monitoring of blood glucose (SMBG) devices1,533 data points
Participants
12 users
Comparison
CGM vs SMBG during flight vs ground cabin pressure
Outcome
Accuracy of CGM vs SMBG: Spearman's correlations and mean absolute relative differences (MARDs)
Methodological detailshypobaric flight simulation, two simulation protocols
  • Hypobaric flight simulation
  • Two simulation protocols

Main finding

As reported in the abstract

Agreement of CGM- and SMBG-derived glucose

Spearman's r, Protocol A, flight vs ground

r=0.96 vs r=0.94

Spearman's r, Protocol B, flight vs ground

r=0.85 vs r=0.69

Aggregated CGM MARD, Protocol A, flight vs ground (minimal difference)

11.85% (95% CI 9.78, 13.92) vs 9.08% (95% CI 7.02, 11.14)

Aggregated CGM MARD, Protocol B, flight vs ground (minimal difference)

12.01% (95% CI 3.34, 20.69) vs 12.97% (95% CI 4.30, 21.65)

Original publication

Comparison of continuous glucose monitoring with self-monitoring of blood glucose in type 1 diabetes in the changing atmospheric pressures in aviation: a hypobaric flight simulation. Diabetologia. Published 31 January 2025. doi:10.1007/s00125-025-06364-z

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.