Diabetes technology

Effects of atmospheric pressure change during flight on insulin pump delivery and glycaemic control of pilots with insulin-treated diabetes: an in vitro simulation and a retrospective observational real-world study

DiabetologiaPublished In vitro simulation and retrospective cohort study

Authors' conclusion

Ambient pressure reduction during simulated flights results in bubble formation and expansion within insulin cartridges. This causes unintended delivery of small insulin doses independent of pre-determined delivery rates and represents the maximum amount of insulin that could be delivered and retracted. However, in vivo, pilots using CSII in-flight did not experience a fall in blood glucose or episodes of hypoglycaemia during these atmospheric pressure changes and the use of insulin pumps can be endorsed in view of their clinical benefits.

Study context

Population
Commercial and private pilots with insulin-treated diabetes granted a medical certificate to fly within the European Union Aviation Safety Agency approved protocol; in vitro insulin pumps in hypobaric chamber de-pressurised to 550 mmHg
Exposure
Continuous subcutaneous insulin infusion (CSII) vs multiple daily injections (MDI) in pilots with insulin-treated diabetes; insulin pump delivery under simulated flight pressure changes
  • CSIIn = 7
Outcome
Insulin delivery vs ground-level performance (in vitro, full cartridges)
Methodological detailshypobaric chamber simulation, 7.5-year retrospective cohort, real-world observational
  • Hypobaric chamber simulation
  • 7.5-year retrospective cohort
  • Real-world observational

Main finding

As reported in the abstract

Over-delivery during 20 min ascent

0.60 U

Under-delivery during descent

0.51 U

Excess insulin delivered during emergency rapid decompression

5.6 U

Observational data: associations, not effects.

Further findings

Insulin delivery during simulated flight (in vitro, hypobaric chamber 550 mmHg), full cartridges vs ground-level performance

U

Over-delivered during 20 min ascent
0.60
Under-delivered during descent
0.51
Excess insulin delivered during emergency rapid decompression
5.6
  • In vitro simulation

In-flight self-monitored blood glucose (SMBG) in pilots using CSII

Pilots using CSII
7
SMBG values recorded during 2345 h of flying across 1081 flights
4656
Values outside acceptable safe range (5.0-15.0 mmol/l [90-270 mg/dl])
33 (0.7%)
No clinically significant fall in median SMBG after aircraft ascent
  • Retrospective cohort, 7.5 years

CSII vs MDI pilots: SMBG values

Values within acceptable range, CSII vs MDI
99.3% vs 97.5%
Values in low red range, CSII vs MDI
0.02% vs 0.1%
In-flight out-of-range values, CSII vs MDI
0.2% vs 1.3%
CSII pilots maintained stricter glycaemic control during flight

Safety

In-flight hypoglycaemia (<4.0 mmol/l [<72 mg/dl]) in pilots using CSII

No in-flight SMBG values within the hypoglycaemic range (<4.0 mmol/l [<72 mg/dl])

Original publication

Effects of atmospheric pressure change during flight on insulin pump delivery and glycaemic control of pilots with insulin-treated diabetes: an in vitro simulation and a retrospective observational real-world study. Diabetologia. Published 4 November 2024. doi:10.1007/s00125-024-06295-1

Composed by AI from the published abstract and checked against it. How Briefs are made →