Type 1 diabetes

Clinical Evaluation of a Novel Insulin Immunosensor

J Diabetes Sci TechnolPublished Diagnostic accuracy study

Authors' conclusion

This proposed technology and preliminary feasibility evaluation show encouraging results for near real-time evaluation of insulin levels, with the potential to improve diabetes management. Real-time measurements of insulin provide person-specific insulin dynamics that could be used to make more informed decisions regarding insulin dosing, thus helping to prevent hypoglycemia and improve diabetes outcomes.

Study context

Population
Adults with T1D after preprandial insulin injectioncompleted the study
Participants
9 adults
Comparison
Insulin immunosensor vs laboratory insulin enzyme-linked immunosorbant assay (ELISA)
Outcome
Accuracy of the insulin immunosensor: concordance of change from baseline in insulin levels with ELISA
Methodological detailsfirst clinical evaluation, preliminary feasibility evaluation
  • First clinical evaluation
  • Preliminary feasibility evaluation

Performance

As reported in the abstract

Concordant paired immunosensor-reference values
Test or modelValueConcordant paired values vs ELISA
Immunosensor70%70%

Immunosensor

Value70%
Concordant paired values vs ELISA70%

Original publication

Clinical Evaluation of a Novel Insulin Immunosensor. J Diabetes Sci Technol. Published 4 February 2022. doi:10.1177/19322968221074406