Diabetes technology

An Automated Insulin Delivery System with Automatic Meal Bolus Based on a Hand-Gesturing Algorithm

Diabetes Technol TherPublished Feasibility trial

Authors' conclusion

Preliminary findings show that a HFCL system was safe and maintained overall glycemic control, similar to that observed with traditional CC and manual meal bolusing.

Study context

Population
Adults with T1D, 18–75 years of age
Participants
17 adults
Comparison
Hands-free closed-loop (HFCL) with gesture-based automatic meal bolus vs traditional carbohydrate counting with meal announcement
Outcome
Overall time in range (70-180 mg/dL)
Methodological detailssequential two-phase, home then supervised hotel setting, Same eight test meals (six solid, two liquid) at the same time and day of the week in both phases
  • Sequential two-phase
  • Home then supervised hotel setting
  • Same eight test meals (six solid, two liquid) at the same time and day of the week in both phases

Main finding

As reported in the abstract

P = 0.08

Further findings

Average

OutcomeHome-stayHotel-stayP value
Average time at <70 mg/dL3.1%3.0%P = 0.9144
Average time at >180 mg/dL13.5%16.3%P = 0.1046

Postprandial glycemia after test meals

Postprandial glycemia following low-carb test meals was similar between the two phases
The system's ability to accommodate high-carb meals was somewhat limited

Safety

Episodes of severe hypoglycemia or DKA
0

Original publication

An Automated Insulin Delivery System with Automatic Meal Bolus Based on a Hand-Gesturing Algorithm. Diabetes Technol Ther. Published 12 March 2024. doi:10.1089/dia.2023.0529

Composed by AI from the published abstract and checked against it. How Briefs are made →