Gestational diabetes & diabetes in pregnancy

Accuracy of intermittently scanned continuous glucose monitoring during caesarean delivery in pregnant women with insulin-treated diabetes

Diabetes Res Clin PractPublished Prospective diagnostic accuracy study

Authors' conclusion

The accuracy of isCGM during caesarean delivery was good, particularly for glucose values between 70 and 110 mg/dL, when CG confirmation could be safely avoided.

Study context

Population
Pregnant women with insulin-treated diabetes during caesarean delivery68 paired isCGM-CG data
Participants
19 women
Comparison
Intermittently scanned CGM (isCGM) vs simultaneously measured capillary glucose
Outcome
Intrapartum accuracy of isCGM vs CG (MARD and Clarke Error Grid analysis)
Follow-up
During caesarean delivery
Methodological detailspaired measurements
  • Paired measurements

Performance

As reported in the abstract

isCGM accuracy vs simultaneous capillary glucose (MARD, Clarke Error Grid)

Overall MARD

9.28 %

Further findings

OutcomeisCGM
Impact on intrapartum management: isCGM-CG pairs in the same glycemic range (same management)68 %
isCGM >110 mg/dL for discordant result at CG (ROC curve)
Sensitivity100 %
AUC0.859
Pp < 0.001
Impact on intrapartum management: isCGM-CG pairs in the same glycemic range (same management) — All discordant data identified by checking capillary glucose (CG) when isCGM >110 mg/dL or <70 mg/dL (chi-square 21.76)
p < 0.001

Original publication

Accuracy of intermittently scanned continuous glucose monitoring during caesarean delivery in pregnant women with insulin-treated diabetes. Diabetes Res Clin Pract. Published 11 March 2024. doi:10.1016/j.diabres.2024.111611