Accuracy of intermittently scanned continuous glucose monitoring during caesarean delivery in pregnant women with insulin-treated diabetes
Authors' conclusion
The accuracy of isCGM during caesarean delivery was good, particularly for glucose values between 70 and 110 mg/dL, when CG confirmation could be safely avoided.
Study context
- Population
- Pregnant women with insulin-treated diabetes during caesarean delivery68 paired isCGM-CG data
- Participants
- 19 women
- Comparison
- Intermittently scanned CGM (isCGM) vs simultaneously measured capillary glucose
- Outcome
- Intrapartum accuracy of isCGM vs CG (MARD and Clarke Error Grid analysis)
- Follow-up
- During caesarean delivery
Methodological detailspaired measurements
- Paired measurements
Performance
As reported in the abstract
isCGM accuracy vs simultaneous capillary glucose (MARD, Clarke Error Grid)
Overall MARD
9.28 %
Further findings
| Outcome | isCGM |
|---|---|
| Impact on intrapartum management: isCGM-CG pairs in the same glycemic range (same management) | 68 % |
| isCGM >110 mg/dL for discordant result at CG (ROC curve) | |
| Sensitivity | 100 % |
| AUC | 0.859 |
| P | p < 0.001 |
- Impact on intrapartum management: isCGM-CG pairs in the same glycemic range (same management) — All discordant data identified by checking capillary glucose (CG) when isCGM >110 mg/dL or <70 mg/dL (chi-square 21.76)
- p < 0.001
Original publication
Accuracy of intermittently scanned continuous glucose monitoring during caesarean delivery in pregnant women with insulin-treated diabetes. Diabetes Res Clin Pract. Published 11 March 2024. doi:10. 1016/ j. diabres. 2024. 111611
Composed by AI from the published abstract and checked against it. How Briefs are made →