Type 1 diabetes

Accuracy of FreeStyle Libre 3 During Moderate-Intensity Continuous Aerobic Exercise at Different Phases of the Menstrual Cycle in Females with Type 1 Diabetes

Diabetes Technol TherPublished Accuracy study

Authors' conclusion

The findings suggest that the accuracy of FSL3 is not affected by CONT, showing good accuracy levels in both phases of the menstrual cycle.

Study context

Population
Females with T1D; mean age 32.2 ± 6.1 years, mean diabetes duration 16.4 ± 8.4 years
Participants
26 females
Comparison
FSL3 vs YSI 2500 during CONT in follicular vs luteal phase
Outcome
Accuracy of FSL3 vs YSI 2500 (MARD) during continuous moderate-intensity aerobic exercise
Methodological detailsprospectively registered, Cycle ergometer sessions; glucose measured every 10 min by FSL3 and YSI 2500, from 20 min before to 20 min after CONT
  • Prospectively registered
  • Cycle ergometer sessions; glucose measured every 10 min by FSL3 and YSI 2500, from 20 min before to 20 min after CONT

Performance

As reported in the abstract

MARD

Follicular

16.06%

Luteal

16.43

FSL3 differed significantly from YSI in both phases; FSL3 higher
about 16 mg/dL

Further findings

MARD pre- vs postexercise, follicular phase

Pre-exercise
14.21%
Postexercise
17.63%
Exercise did not affect mean absolute relative differences (MARDs)

MARD pre- vs postexercise, luteal phase

Pre-exercise
14.95%
Postexercise
17.71%

Original publication

Accuracy of FreeStyle Libre 3 During Moderate-Intensity Continuous Aerobic Exercise at Different Phases of the Menstrual Cycle in Females with Type 1 Diabetes. Diabetes Technol Ther. Published 17 April 2025. doi:10.1089/dia.2024.0558

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.