Ultrasound Detected Subcutaneous Changes in a Pediatric Cohort After Initiation of a New Insulin Pump or Glucose Sensor

Diabetes Technol TherPublished Prospective cohort study

Authors' conclusion

There is large variation in the distance from the skin surface to the muscle fascia and many diabetes devices reach even deeper. Hyperechogenicity and vascularization increased significantly over time at CSII sites, but not CGM sites.

Study context

Population
Children and adolescents during the first year after initiation of a new diabetes device (continuous subcutaneous insulin infusion (CSII) or continuous glucose monitoring (CGM))
Participants
161 children
Exposure
CSII vs CGM
Outcome
Subcutaneous tissue changes (echogenicity, vascularization, skin-to-muscle distance) at CSII and CGM sites by US, and association with HbA1c
Follow-up
First year
Methodological detailsultrasound assessment
  • Ultrasound assessment

Main finding

As reported in the abstract

Subcutaneous changes at device sites
Mean distance for boys, abdomen (all ages)
4.5-6.5 mm
Mean distance for boys, upper arm (all ages)
5-6.9 mm
Hyperechogenicity at CGM sites after 12 months
4.3%
Hyperechogenicity at CSII sites increased over time
41.2% to 69.3%; P < 0.001
Vascularization at CSII sites increased over time
2% to 16%; P = 0.009

Observational data: associations, not effects.

Further findings

Hyperechogenicity as predictor of elevated HbA1c

Hyperechogenicity in the subcutis was not a predictor of elevated HbA1c
P = 0.11

Original publication

Ultrasound Detected Subcutaneous Changes in a Pediatric Cohort After Initiation of a New Insulin Pump or Glucose Sensor. Diabetes Technol Ther. Published 6 June 2023. doi:10.1089/dia.2023.0137

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.