Diabetes technology

A New Index of Insulin Sensitivity from Glucose Sensor and Insulin Pump Data: In Silico and In Vivo Validation in Youths with Type 1 Diabetes

Diabetes Technol TherPublished Validation study

Authors' conclusion

SISP is suitable for estimating SI in both closed- and open-loop settings, provided that the subject wears a CGM sensor and a subcutaneous insulin pump.

Study context

Population
Virtual adults (UVa/Padova T1D Simulator) and youths with T1D; two consecutive meals (breakfast and lunch, at 7 and 11 am) with 70 g carbohydrate each
Participants
100 vs 10
Comparison
SISP vs oral glucose minimal model sensitivity (SI)
  • Simulatedn = 100
  • Realn = 10
Outcome
SISP estimation precision and correlation with SIMM
Methodological detailsin silico and in vivo, simulated and real datasets, validation against gold standard SIMM
  • In silico and in vivo
  • Simulated and real datasets
  • Validation against gold standard SIMM

Performance

As reported in the abstract

Test or modelValueMeals with good precision (median CV <20%)R (SISP vs SIMM)
Real100%100%0.83
Simulated91%91%0.82

Value

Real100%
Simulated91%

Meals with good precision (median CV <20%)

Real100%
Simulated91%

R (SISP vs SIMM)

Real0.83
Simulated0.82
SISP and SIMM highly correlated, P < 0.001
R = 0.82 and R = 0.83

Original publication

A New Index of Insulin Sensitivity from Glucose Sensor and Insulin Pump Data: In Silico and In Vivo Validation in Youths with Type 1 Diabetes. Diabetes Technol Ther. Published 17 February 2023. doi:10.1089/dia.2022.0397

Composed by AI from the published abstract and checked against it. How Briefs are made →