Thyroid disease

A Preliminary Evaluation of the Diagnostic Performance of a Smartphone-Based Machine Learning-Assisted System for Evaluation of Clinical Activity Score in Digital Images of Thyroid-Associated Orbitopathy

ThyroidPublished Diagnostic accuracy study

Authors' conclusion

Our smartphone-based ML-assisted system shows reasonable accuracy in detecting active TAO, comparable with oculoplastic specialists and outperforming residents and general ophthalmologists.

Study context

Population
Patients with thyroid-associated orbitopathy (TAO); reference clinical activity score (CAS) by consensus of three oculoplastic specialists28 with active TAO
Participants
100 patients
Comparison
Smartphone-based machine learning (ML)-assisted system vs ophthalmologists with varying clinical experience for detecting active TAO (CAS ≥3)
  • Ophthalmology residentsn = 3
  • General ophthalmologistsn = 3
  • Oculoplastic specialistsn = 3
Outcome
Detection of active TAO (CAS ≥3): sensitivity, specificity and F1 score
Methodological detailspreliminary evaluation
  • Preliminary evaluation

Performance

As reported in the abstract

Detection of active TAO (CAS ≥3) vs reference CAS
Test or modelSensitivity (top 3 smartphones)SensitivitySpecificity (top 3 smartphones)SpecificityF1 score (top 3 smartphones)F1 score
Smartphone-based system74.5%— not stated in the abstract84.8%— not stated in the abstract0.70— not stated in the abstract
Ophthalmology residents— not stated in the abstract69.1%— not stated in the abstract55.1%— not stated in the abstract0.46
General ophthalmologists— not stated in the abstract61.9%— not stated in the abstract79.6%— not stated in the abstract0.55
Oculoplastic specialists— not stated in the abstract73.8%— not stated in the abstract90.7%— not stated in the abstract0.75

Sensitivity (top 3 smartphones)

Smartphone-based system74.5%
Ophthalmology residents— not stated in the abstract
General ophthalmologists— not stated in the abstract
Oculoplastic specialists— not stated in the abstract

Sensitivity

Smartphone-based system— not stated in the abstract
Ophthalmology residents69.1%
General ophthalmologists61.9%
Oculoplastic specialists73.8%

Specificity (top 3 smartphones)

Smartphone-based system84.8%
Ophthalmology residents— not stated in the abstract
General ophthalmologists— not stated in the abstract
Oculoplastic specialists— not stated in the abstract

Specificity

Smartphone-based system— not stated in the abstract
Ophthalmology residents55.1%
General ophthalmologists79.6%
Oculoplastic specialists90.7%

F1 score (top 3 smartphones)

Smartphone-based system0.70
Ophthalmology residents— not stated in the abstract
General ophthalmologists— not stated in the abstract
Oculoplastic specialists— not stated in the abstract

F1 score

Smartphone-based system— not stated in the abstract
Ophthalmology residents0.46
General ophthalmologists0.55
Oculoplastic specialists0.75

— not stated in the abstract

Average sensitivity on images from all 6 smartphones
71.4%
Average specificity on images from all 6 smartphones
81.6%
Average F1 score on images from all 6 smartphones
0.66
F1 score range by smartphone model
0.59-0.72

Further findings

OutcomeValues
Predicted CAS within 1 point of reference CASSmartphone-based ML system 90.7%

Original publication

A Preliminary Evaluation of the Diagnostic Performance of a Smartphone-Based Machine Learning-Assisted System for Evaluation of Clinical Activity Score in Digital Images of Thyroid-Associated Orbitopathy. Thyroid. Published 2 May 2024. doi:10.1089/thy.2023.0621

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.