Diabetes technology

Performance of an Automated Insulin Delivery System in People Living With Type 2 Diabetes and Insulin Resistance: First Real-World Evidence in 26,427 Users

J Diabetes Sci TechnolPublished Retrospective real-world analysis

Authors' conclusion

The use of this AID is associated with effective therapy outcomes, as indicated by over 70% TIR, and appears to be safe, as demonstrated by a low TBR in a large cohort of real-life users with self-reported T2D and high or low TDD.

Study context

Population
MiniMed 780G automated insulin delivery (AID) users with T2D phenotypes, data uploaded to CareLink Personal (January 2020 to April 2024)
Participants
18,466 / 10,795 / 2,834 / 7,961
Exposure
Cohorts of MM780G AID users with T2D phenotypes by insulin requirement
  • Cohort An = 18,466
  • Cohort Bn = 10,795
  • Cohort Cn = 2,834
  • Cohort Dn = 7,961
Outcome
TIR
Follow-up
Post-AID

Main finding

As reported in the abstract

Observational data: associations, not effects.

Further findings

Time in range (TIR) in cohort C using recommended optimal settings (GT 100 mg/dL, AIT 2 hours)
78.7 · ± 10.8 · %

TIR post-AID vs pre-AID

All cohorts increased ≥10% post-AID compared with pre-AID

Safety

Mean time below range (TBR) <70 mg/dL in all cohorts
≤1%

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

Performance of an Automated Insulin Delivery System in People Living With Type 2 Diabetes and Insulin Resistance: First Real-World Evidence in 26,427 Users. J Diabetes Sci Technol. Published 12 March 2025. doi:10.1177/19322968251318373

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.