Diabetes technology

The Performance of Freestyle Libre Pro Flash Continuous Glucose Monitoring in Hospitalized Patients Treated with an Intravenous Insulin Infusion for Acute Prednisolone-Induced Hyperglycemia

Diabetes Technol TherPublished Prospective study

Main finding · Paired glucose differenceResult — go to the resultMean difference 1.5 mmol/L

Study context

Population
Hospitalized adults with prednisolone-associated hyperglycemia on IV insulin infusion adjusted hourly using flash glucose monitoring
Exposure
Flash glucose monitoring vs point of care (POC) glucose
Outcome
Difference in paired POC and flash glucose measurements and risk of severe hyper- or hypoglycemia (Clarke error grid analysis)

Main finding

Paired glucose difference

Mean difference

1.5 mmol/L

P < 0.001

By group

POC

9.1

mean · ± 4.1 · mmol/L

As reported in the abstract

Glucose measurements within Zone A and B on error grid analysis
97.8%

Observational data: associations, not effects.

Original publication

The Performance of Freestyle Libre Pro Flash Continuous Glucose Monitoring in Hospitalized Patients Treated with an Intravenous Insulin Infusion for Acute Prednisolone-Induced Hyperglycemia. Diabetes Technol Ther. Published 5 December 2023. doi:10.1089/dia.2023.0360

Composed by AI from the published abstract and checked against it. How Briefs are made →