Thyroid disease

How low can we (reliably) go? A method comparison of thyroid-stimulating hormone assays with a focus on low concentrations

Eur Thyroid JPublished Method comparison study

Authors' conclusion

We established the standardization differences of frequently used TSH assays for the total and low concentration ranges. Based on the proportional bias and the imprecision, this effect seems to have limited clinical consequences for the low TSH concentration range. Nevertheless, as guidelines mention absolute TSH values to guide clinical decision-making, caution must be applied when interpreting values close to these cut-offs.

Study context

Population
Samples covering a wide range of thyroid-stimulating hormone (TSH) concentrations (<0.01 to 120 mIU/L), oversampled in the lower range (<0.4 mIU/L)
Participants
60 samples
Comparison
Three TSH immunoassays
Outcome
Standardization differences (bias) between TSH immunoassays

Main finding

As reported in the abstract

Bias, total range

7-16%

Bias, low range

1-14%

Different classification of TSH concentrations <0.40 mIU/L measured over last 6 months, at cut-off TSH = 0.1 mIU/L

1.5%

Further findings

Imprecision (CV from duplicate measurements, 20 samples)

Imprecision of the immunoassays
1.6-5.5%
Could lead to a reclassification similar to that from bias between immunoassays

Original publication

How low can we (reliably) go? A method comparison of thyroid-stimulating hormone assays with a focus on low concentrations. Eur Thyroid J. Published 14 September 2023. doi:10.1530/etj-23-0123