Diabetes technology

Glycemic Concordance of Simplera Sync ™ and Instinct ™ Sensors in MiniMed ™ 780G System Users: A Randomized Crossover Feasibility Study

Diabetes Technol TherPublished Randomised crossover trial

Primary result · TIR (70-180 mg/dL)Result — go to the resultMean difference 1.5 % 95% CI 0.18–2.79

Authors' conclusion

This randomized controlled crossover trial demonstrates that both Simplera Sync and Instinct sensors meet international consensus targets for glycemic control when used with the MM780G system, and that the sensors can be interchanged without meaningful differences in glycemic control.

Study context

Population
Adults with T1D using the MiniMed 780G system23 completed
Participants
24 adults
Comparison
Simplera Sync vs Instinct sensor with MM780G
Primary endpoint
Noninferiority in time in range (TIR) (70-180 mg/dL), noninferiority margin 7.5%
Follow-up
6-week sensor periods
Methodological detailssingle-center, two-arm, noninferiority, feasibility
  • Single-center
  • Two-arm
  • Noninferiority
  • Feasibility

Primary result

TIR (70-180 mg/dL) · 6-week period

Mean difference

1.5 %

95% CI 0.18–2.79

By group

Simplera Syncvs Instinct
mean difference 1.5 % (95% CI 0.18–2.79)

Non-inferiority margin 7.5 as printed, not drawn

As reported in the abstract

Secondary outcomes

HbA1c at end of sensor period · end of 6-week period

Safety

Serious adverse events related to sensor use
0

Original publication

Glycemic Concordance of Simplera Sync ™ and Instinct ™ Sensors in MiniMed ™ 780G System Users: A Randomized Crossover Feasibility Study. Diabetes Technol Ther. Published 21 May 2026. doi:10.1177/15209156261449884

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.