Diabetes technology

First Outpatient Clinical Trial of a Full Closed-Loop Artificial Pancreas System in South America

J Diabetes Sci TechnolPublished Outpatient clinical trial

Authors' conclusion

The results obtained demonstrated the feasibility of outpatient full CL regulation of glucose levels involving the ARG algorithm and the InsuMate platform.

Study context

Population
Adults with meals of 38 to 120 g CHO and daily mild exercise (3–5 km walks), outpatient setting
Participants
5 adults
Comparison
Automatic regulation of glucose (ARG) algorithm full closed-loop vs open-loop control
Outcome
Glycaemic control
Follow-up
72 hours
Methodological detailsphase 3, crossover of 72 h closed-loop and 72 h open-loop control, Own artificial pancreas (AP) and remote monitoring platform (InsuMate) developed from open-source resources
  • Phase 3
  • Crossover of 72 h closed-loop and 72 h open-loop control
  • Own AP and remote monitoring platform (InsuMate) developed from open-source resources

Main finding

As reported in the abstract

Time in hyperglycemia, open-loop (OL) vs closed-loop (CL)

52.2% [16.3%] OL vs 48.0% [15.4%] CL

TIR, OL vs CL

46.9% [15.6%] OL vs 50.9% [14.4%] CL

Mean glucose, OL vs CL

188.9 [25.5] mg/dl OL vs 186.2 [24.7] mg/dl CL

Average time in CL achieved by InsuMate

more than 95%

Safety

No severe hyperglycemia or hypoglycemia episodes during the trial

Original publication

First Outpatient Clinical Trial of a Full Closed-Loop Artificial Pancreas System in South America. J Diabetes Sci Technol. Published 12 May 2022. doi:10.1177/19322968221096162