Feasibility study of a combined insulin‐delivery and glucose sensor system worn over 7 days in persons with type 1 diabetes

Diabetes Obes MetabPublished Feasibility study

Authors' conclusion

This human feasibility study supports the potential of CGM-IS use for up to 7 days with a mean survival of over 6 days. However, our data indicate that the durability of insulin delivery and sensor refinement are required prior to commercialisation.

Study context

Population
Adults with T1D using insulin pumps and continuous glucose monitoring (CGM); 7 days with three meal test days and four free living days
Participants
40 adults
Outcome
Sensor accuracy (MARD) and infusion set function (TIR, mean sensor glucose, total daily dose of insulin)
Follow-up
7-day wear
Methodological detailssingle centre, single arm, non-randomised
  • Single centre
  • Single arm
  • Non-randomised

Main finding

As reported in the abstract

Sensor mean average relative difference (MARD), Day 1
16.4%
Sensor MARD, Day 4
14.8%
Sensor MARD, Day 7
17.7%
Readings meeting 15/15 criterion for blood glucose (BG) readings <70.2 mg/dL
60.0%
Total daily dose (TDD) increase at Days 5-6
7.3 units/day (16%)
Mean BG increase at Days 5-6
16.2 mg/dL (10.2%)
Time in range (TIR) fall at Days 5-6
−7.8% (11.2%)

Further findings

Device survival and wear time

Devices lasting >4 days
92.1%
Mean wear time
6.3 ± 1.4 days

Safety

Tolerability

The device was well tolerated

Original publication

Feasibility study of a combined insulin‐delivery and glucose sensor system worn over 7 days in persons with type 1 diabetes. Diabetes Obes Metab. Published 25 November 2025. doi:10.1111/dom.70331