Diabetes technology

Evaluation of a Fourth-Generation Subcutaneous Real-Time Continuous Glucose Monitor (CGM) in Individuals With Diabetes on Peritoneal Dialysis

Diabetes CarePublished Diagnostic accuracy study

Authors' conclusion

We showed satisfactory performance of a real-time CGM sensor in PD patients with diabetes, supporting future use to facilitate treatment decisions.

Study context

Population
Participants with T2D on continuous ambulatory peritoneal dialysis (PD)
Participants
30 participants
Comparison
CGM vs Yellow Springs Instrument (YSI) venous glucose
  • Guardian Sensor 3n = 30
Outcome
Accuracy of CGM vs YSI venous glucose
Follow-up
8-h in-clinic session with glucose challenge
Methodological details14-day sensor wear, 8-h in-clinic session with glucose challenge
  • 14-day sensor wear
  • 8-h in-clinic session with glucose challenge

Performance

As reported in the abstract

Mean absolute relative difference (MARD) (95% CI 9.6, 11.7), 941 CGM-YSI matched pairs

10.4%

Readings within %15/15 of YSI values, full glycemic range

81.3%

Consensus error grid: sensor values in zones A and B

99.9%

Further findings

No correlations between pH, uremia, hydration status and MARD

Original publication

Evaluation of a Fourth-Generation Subcutaneous Real-Time Continuous Glucose Monitor (CGM) in Individuals With Diabetes on Peritoneal Dialysis. Diabetes Care. Published 12 April 2023. doi:10.2337/dc22-2348