Diabetes technology

Evaluating the Impact of Applying Personal Glucose Targets in a Closed-Loop System for People With Type 1 Diabetes

J Diabetes Sci TechnolPublished Post-hoc analysis of five randomised controlled trials

Authors' conclusion

Discrete study populations showed differences in glucose control when applying similar personal targets.

Study context

Population
Participants with T1D using the CamAPS FX hybrid closed-loop app in five RCTs5 RCTs
Comparison
Personal glucose targets vs default target bin (5.5-6.0 mmol/L) in CamAPS FX hybrid closed-loop

Primary result

As reported in the abstract

TIR (3.9-10.0 mmol/L [70-180 mg/dL])
Each group vs default target binMean difference, percentage points (95% CI)P value
6.5-7.0 mmol/L
−3.2(−5.3 to −1.2)
P < 0.001
7.0-7.5 mmol/L
−10.8(−14.1 to −7.6)
P < 0.001
Days with a target outside the default target bin
20.3%

Further findings

OutcomeEstimateP value
Time <3.9 mmol/L [<70 mg/dL]6.5-7.0 mmol/L vs Default target bin−1.85 percentage points (95% CI −2.37 to −1.34)P < 0.001
Time <3.9 mmol/L [<70 mg/dL]7.0-7.5 mmol/L vs Default target bin−2.68 percentage points (95% CI −3.49 to −1.86)P < 0.001

Original publication

Evaluating the Impact of Applying Personal Glucose Targets in a Closed-Loop System for People With Type 1 Diabetes. J Diabetes Sci Technol. Published 20 December 2022. doi:10.1177/19322968221145184