Defining the Functional Sensitivity for the Siemens Atellica Calcitonin Assay: Insight From a Single-Center Study

Endocr PractPublished Retrospective diagnostic accuracy study

Authors' conclusion

Our findings indicate the potential suitability of a 5 pg/mL calcitonin cutoff on the Siemens Atellica platform for evaluating tumor persistence or recurrence in post-thyroidectomy patients in our institution. However, individual laboratories should establish their own clinical cutoff value when evaluating calcitonin levels for monitoring tumor recurrence post-thyroidectomy.

Study context

Population
Patients ≥3 months post-total thyroidectomy for MTC with calcitonin measured on Siemens Atellica, Sept 2022–Aug 2023; imaging within 6 months of calcitonin report56 samples
Participants
40 patients
Comparison
Calcitonin cutoffs 1.89 pg/mL vs 5.00 pg/mL on Siemens Atellica assay
Outcome
Diagnostic accuracy of calcitonin cutoffs for persistent or recurrent MTC confirmed by imaging
Methodological detailssingle-centre
  • Single-centre

Performance

As reported in the abstract

Persistent or recurrent MTC confirmed by imaging
StrategySensitivitySpecificity
1.89 pg/mL cutoff43%67%
5.00 pg/mL cutoff0%100%

Sensitivity

1.89 pg/mL cutoff43%
5.00 pg/mL cutoff0%

Specificity

1.89 pg/mL cutoff67%
5.00 pg/mL cutoff100%

Further findings

Assay precision (CV)

CV at 2.94 pg/mL
16.49%
CV at 5.24 pg/mL
8.87%

Original publication

Defining the Functional Sensitivity for the Siemens Atellica Calcitonin Assay: Insight From a Single-Center Study. Endocr Pract. Published 26 August 2024. doi:10.1016/j.eprac.2024.08.007