Diabetes technology

Continuous Glucose Monitoring Using the Dexcom G6 in Cardiac Surgery During the Postoperative Period

Endocr PractPublished Prospective accuracy study

Authors' conclusion

Almost 100% of Dexcom G6Pro CGM and clinical data matching points fell within areas considered as giving clinically correct decisions (zone A) and clinically uncritical decisions (zone B). However, the relatively high mean absolute relative difference precludes its use for both monitoring and treatment in the clinical context. As technology evolves, interstitial glucose monitoring may become an important tool to limit iatrogenic anemia and mitigate glycemic fluctuations.

Study context

Population
Patients in the postoperative cardiac surgery period after intensive care unit (ICU) admission; 71% male, 14% with known diabetes, 66% required IV insulin infusion1,428 clinical readings
Participants
29 patients
Comparison
Dexcom G6Pro CGM vs clinical glucose measurements
Outcome
Accuracy of Dexcom G6Pro CGM vs clinical glucose, all measures post-ICU admission
Methodological detailspilot study, CGM worn until discharge or 10 days postoperatively
  • Pilot study
  • CGM worn until discharge or 10 days postoperatively

Performance

As reported in the abstract

Clarke error grid zone A

53.5%

Clarke error grid zone B

45.9%

Clarke error grid zones D or E

0.6%

n = 5

Mean absolute relative difference

20.6%

Further findings

Accuracy excluding first 24 hours post-ICU admission

Clarke error grid zone A
55.9%
Clarke error grid zone B
43.9%
Clarke error grid zone D
0.2%
Mean absolute relative difference
21.6%

Consensus error grid, subanalysis of 12 patients without >5-minute tap in operating room, post-ICU admission

Zone A
53.9%
Zone B
45.4%
Zone C
0.7%
Similar percentages removing first 24 hours post-ICU admission

Consensus error grid post-ICU admission

Zone A
54%
Zone B
45%
Zone C
0.9%

Consensus error grid excluding first 24 hours

Zone A
56%
Zone B
44%
Zone C
0.4%

Safety

No adverse events were reported

Original publication

Continuous Glucose Monitoring Using the Dexcom G6 in Cardiac Surgery During the Postoperative Period. Endocr Pract. Published 30 April 2024. doi:10.1016/j.eprac.2024.04.015