Gestational diabetes & diabetes in pregnancy
Continuous Glucose Monitoring-Measured Glucose Levels During Oral Glucose Tolerance Testing in Pregnancy
Main finding · CGM vs BG glucose during OGTTResult — go to the resultDifference 6 ± 13 mg/dLAuthors' conclusion
CGM glucose measurements using Dexcom G6 Pro tended to be slightly higher than BG values during an OGTT, leading to more participants who would have been diagnosed with GDM if the BG-based OGTT thresholds were applied to these CGM-measured glucose values.
Study context
- Population
- Pregnant females undergoing diagnostic oral glucose tolerance test (OGTT) while wearing a blinded continuous glucose monitoring (CGM) sensor
- Participants
- 119 females
- Comparison
- CGM-measured glucose vs blood glucose (BG) during OGTT
- Outcome
- CGM vs BG glucose during OGTT
Methodological detailsblinded CGM sensor, Dexcom G6 Pro
- Blinded CGM sensor
- Dexcom G6 Pro
Main finding
As reported in the abstract
Difference
6 ± 13 mg/dL
mg/dL
- Median CGM minus BG glycemic excursion, 75 g OGTT
- 12
- Median CGM minus BG glycemic excursion, 100 g OGTT
- 4
Further findings
Gestational diabetes mellitus (GDM) diagnosis by CGM with BG-based thresholds
- Of 28 diagnosed with GDM by OGTT BG, also diagnosed by CGM
- 24 (86%)
- Of 91 not diagnosed with GDM by BG, also not diagnosed by CGM
- 54 (59%)
Original publication
Continuous Glucose Monitoring-Measured Glucose Levels During Oral Glucose Tolerance Testing in Pregnancy. Diabetes Technol Ther. Published 2 April 2025. doi:10. 1089/ dia. 2024. 0563