Diabetes technology

Comparisons of Fifth-, Sixth-, and Seventh-Generation Continuous Glucose Monitoring Systems

J Diabetes Sci TechnolPublished Review with propensity-score-adjusted comparison of preapproval trial data

Authors' conclusion

Adjusted MARD and %20/20 agreement rates were 9.0%/93.1% (abdomen-placed G5), 9.9%/92.3% (abdomen-placed G6), 9.1%/93.2% (abdomen-placed G7), and 8.2%/95.3% (arm-placed G7). Task analysis favored G7 over earlier systems. Favorable clinical outcomes such as hemoglobin A1c reduction and hypoglycemia avoidance seen with G5 and G6 are anticipated with G7 use.

Study context

Comparison
Dexcom G5 vs G6 vs G7 continuous glucose monitoring (CGM) systems
Outcome
Adjusted MARD vs YSI and %20/20 agreement rate
Methodological detailspropensity score adjustments, formal task analysis
  • Propensity score adjustments
  • Formal task analysis

Main finding

As reported in the abstract

Values

%20/20

Further findings

Ease-of-use (task analysis)

Task analysis favored G7 over earlier systems

Original publication

Comparisons of Fifth-, Sixth-, and Seventh-Generation Continuous Glucose Monitoring Systems. J Diabetes Sci Technol. Published 13 June 2022. doi:10.1177/19322968221099879