Diabetes technology

Clinical Validation on Healthy Humans of a Portable Non-Invasive Continuous Glucose Monitor Based on Transdermal Band-Pass Raman Spectroscopy

J Diabetes Sci TechnolPublished Clinical validation study

Authors' conclusion

These results demonstrate that the BPR-CGM technology can achieve clinical-grade accuracy comparable to invasive systems

Study context

Population
Healthy humans undergoing a standard OGTT
Participants
6 participants
Comparison
Non-invasive BPR-CGM vs standard blood glucometer and two commercial needle-based continuous glucose monitoring (CGMs)
Outcome
Mean absolute relative difference vs reference blood glucose during standard OGTT
Methodological detailsnon-invasive, oral glucose tolerance test
  • Non-invasive
  • Oral glucose tolerance test

Performance

As reported in the abstract

Further findings

Parkes (consensus) error grid

All BPR-CGM readings within clinically acceptable zones A and B

Glucose trends

BPR-CGM trends closely matched reference blood glucose without algorithmic lag compensation

Safety

Adverse skin reactions · following the trial

No adverse skin reactions observed

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

Clinical Validation on Healthy Humans of a Portable Non-Invasive Continuous Glucose Monitor Based on Transdermal Band-Pass Raman Spectroscopy. J Diabetes Sci Technol. Published 30 April 2026. doi:10.1177/19322968261445056

Composed by AI from the published abstract and checked against it. How Briefs are made →