Thyroid disease

Assessment of the Diagnostic Performance of a Commercially Available Artificial Intelligence Algorithm for Risk Stratification of Thyroid Nodules on Ultrasound

ThyroidPublished Diagnostic accuracy study

Authors' conclusion

Koios DS demonstrated standalone performance similar to radiologists, though with lower sensitivity and higher specificity. Performance was best when radiologist interpretations were combined with the AI-adapter component, with improved specificity and reduced unnecessary FNA recommendations.

Study context

Population
Consecutive thyroid ultrasound (US) with ≥1 nodule over one year; ground truth by surgical pathology, fine needle aspiration (FNA), or 3-year follow-up US showing stability649 nodules
Participants
347 participants
Comparison
Food and Drug Administration (FDA)-cleared artificial intelligence (AI) tool (Koios DS Thyroid) with or without AI-adapter vs clinical radiology report, alone or combined with AI-adapter
Outcome
Performance of final recommendations (FNA or no FNA) vs ground truth: sensitivity, specificity, ROC analysis
Methodological detailsretrospective, consecutive, two centres (Duke University Hospital and affiliate community hospital)
  • Retrospective
  • Consecutive
  • Two centres (Duke University Hospital and affiliate community hospital)

Performance

As reported in the abstract

Diagnostic performance of final FNA recommendation vs ground truth
Test or modelAUCSensitivitySpecificity
Koios with AI-adapter0.70 (CI 0.60-0.81)0.69 (CI: 0.50-0.83)0.63 (CI: 0.59-0.67)
Radiologists0.70 (CI 0.60-0.79)0.81 (CI: 0.61, 0.92)0.43 (CI: 0.38-0.48)
Radiology report + AI-adapter0.76 (CI: 0.67-0.85)0.810.53 (CI: 0.49-0.58)

AUC

Koios with AI-adapter0.70 (CI 0.60-0.81)
Radiologists0.70 (CI 0.60-0.79)
Radiology report + AI-adapter0.76 (CI: 0.67-0.85)

Sensitivity

Koios with AI-adapter0.69 (CI: 0.50-0.83)
Radiologists0.81 (CI: 0.61, 0.92)
Radiology report + AI-adapter0.81

Specificity

Koios with AI-adapter0.63 (CI: 0.59-0.67)
Radiologists0.43 (CI: 0.38-0.48)
Radiology report + AI-adapter0.53 (CI: 0.49-0.58)
  • Radiology report + AI-adapter vs Radiologists: McNemar's test p < 0.001 — specificity improved from 0.43 to 0.53
  • Radiology report + AI-adapter vs Radiologists: P = 1 — unchanged sensitivity
Fewer FNA recommendations with radiology report + AI-adapter
17%

Original publication

Assessment of the Diagnostic Performance of a Commercially Available Artificial Intelligence Algorithm for Risk Stratification of Thyroid Nodules on Ultrasound. Thyroid. Published 15 October 2024. doi:10.1089/thy.2024.0410

Composed by AI from the published abstract and checked against it. How Briefs are made →