Diabetes technology

Assessment of the Impact of Subcutaneous Catheter Change on Glucose Control in Patients with Type 1 Diabetes Treated by Insulin Pump in Open- and Closed-Loop Modes

Diabetes Technol TherPublished Retrospective analysis

Authors' conclusion

This study shows an increase in time spent in hyperglycemia on the day of the KT change both in OL and AID modes.

Study context

Population
Patients with T1D using Tandem t:slim x2 pump and Dexcom G6 sensor for 20 days in open-loop (OL), then as automated insulin delivery (AID); 72 women (54%), age 35.6 ± 15.7 years, body mass index (BMI) 25.2 ± 4.7 kg/m², hemoglobin A1c (HbA1c) 7.5 ± 0.8%1,636 KT changes
Participants
134 patients
Exposure
%TAR180 on day of catheter change (D0) vs next day (D1) and 2 days later (D2), in OL and AID modes
Outcome
Percentage of time with sensor glucose >180 mg/dL (%TAR180)
Methodological detailsadjusted for age, gender, BMI, HbA1c at inclusion and %TAR180 for the 2 h before KT change
  • Adjusted for age, gender, BMI, HbA1c at inclusion and %TAR180 for the 2 h before KT change

Main finding

As reported in the abstract

%TAR180
D0 vs D1 · OL mode
absolute increase +6.9 %; P < 0.0001
D0 vs D2 · OL mode
absolute increase +6.8 %; P < 0.0001
D0 vs D1 · AID mode
absolute increase +4.8 %; P < 0.0001
D0 vs D2 · AID mode
absolute increase +4.2 %; P < 0.0001

Observational data: associations, not effects.

Further findings

%TAR180 in the 2 h before KT change

OL mode
51.3 ± 37.0%
AID mode
33.2 ± 30.0%

Original publication

Assessment of the Impact of Subcutaneous Catheter Change on Glucose Control in Patients with Type 1 Diabetes Treated by Insulin Pump in Open- and Closed-Loop Modes. Diabetes Technol Ther. Published 7 March 2024. doi:10.1089/dia.2023.0568

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.