Accuracy of Continuous Glucose Monitors for Inpatient Diabetes Management

J Diabetes Sci TechnolPublished Diagnostic accuracy study

Authors' conclusion

Despite reduced accuracy compared with the outpatient setting, both versions of CGMs had acceptable safety profiles in the inpatient setting.

Study context

Population
Adult patients with diabetes admitted to non-ICU medical wards
Participants
34 vs 43
Comparison
Continuous glucose monitoring (CGM) version 2 vs CGM version 1, accuracy relative to point-of-care (POC) capillary glucose
  • CGM version 2n = 34
  • CGM version 1n = 43reference
Outcome
CGM accuracy relative to POC (mean absolute relative difference (MARD)) and safety of use (Clarke Error Grid analysis)

Performance

As reported in the abstract

MARD vs POC

CGM version 2

17.7

%

CGM version 1

21.4%

Further findings

Clarke Error Grid: paired values in acceptable zones A and B

Paired values in zones A and B
98.8%

CGM accuracy by POC value and time of sensor wear

CGM accuracy did not change with POC value or with time of sensor wear

Safety

Safety of use (Clarke Error Grid)

Paired values within acceptable zones A and B
98.8%

Original publication

Accuracy of Continuous Glucose Monitors for Inpatient Diabetes Management. J Diabetes Sci Technol. Published 7 February 2022. doi:10.1177/19322968221076562

Composed by AI from the published abstract and checked against it. How Briefs are made →