Accuracy of Continuous Glucose Monitoring in Hemodialysis Patients With Diabetes
Performance · Accuracy of CGM vs blood glucoseResult — go to the resultMean difference (Bland-Altman) ∼+15 mg/dLAuthors' conclusion
In hemodialysis patients with diabetes, although MARD values were higher than traditional optimal analytic performance thresholds, error grids showed nearly all CGM values were in clinically acceptable zones. Further studies are needed to determine whether CGM improves outcomes in hemodialysis patients.
Study context
- Population
- Hospitalized maintenance hemodialysis patients with diabetes, October 2020 to May 2021
- Participants
- 31 patients
- Comparison
- Dexcom G6 continuous glucose monitoring (CGM) vs blood glucose
- Outcome
- Accuracy of CGM vs blood glucose
Methodological detailsprotocolized glucose measurements
- Protocolized glucose measurements
Performance
As reported in the abstract
Mean difference (Bland-Altman)
∼+15 mg/dL
- Pearson correlation, averaged CGM vs blood glucose
- 0.84
- Spearman correlation, averaged CGM vs blood glucose
- 0.79
- Agreement rate (%20/20), overall period
- 48.7%
- Agreement rate (%20/20), hemodialysis periods
- 47.2%
- Agreement rate (%20/20), nonhemodialysis periods
- 50.2%
- Mean ARD (MARD) across all time periods
- ∼20%
- Median ARD, overall period
- 19.4%
- Median ARD, nonhemodialysis periods (slightly lower)
- 18.2%
- Median ARD, hemodialysis periods
- 22.0%
Original publication
Accuracy of Continuous Glucose Monitoring in Hemodialysis Patients With Diabetes. Diabetes Care. Published 30 August 2024. doi:10. 2337/ dc24-0635