Accuracy of an Off-Label Transmitter and Data Manager Paired With an Intermittent Scanned Continuous Glucose Monitor in Adults With Type 1 Diabetes
Authors' conclusion
Measurements recorded by the c-rtCGM were found to be accurate when compared with FSL data only when performing daily c-rtCGM device calibrations. High drops in accuracy and agreement between devices occurred when the c-rtCGM was not calibrated.
Study context
- Population
- Adults with T1D
- Participants
- 21 adults
- Comparison
- Off-label converted real-time CGM (converted real-time continuous glucose monitor (c-rtCGM): MiaoMiao transmitter + xDrip+ app) vs FreeStyle Libre (FSL), calibrated vs uncalibrated
- Outcome
- Accuracy of c-rtCGM vs FSL
Performance
As reported in the abstract
| Outcome | Calibrated | Uncalibrated |
|---|---|---|
| Overall MARD | 12.06% | 17.49% |
| Consensus Error Grid Zone A | 84.71% | — not stated in the abstract |
— not stated in the abstract
Original publication
Accuracy of an Off-Label Transmitter and Data Manager Paired With an Intermittent Scanned Continuous Glucose Monitor in Adults With Type 1 Diabetes. J Diabetes Sci Technol. Published 25 October 2022. doi:10. 1177/ 19322968221133405