Pediatric endocrinology

A Preliminary Analysis of Thyrotropin Measurement from Finger Stick Dried Blood Spot with an Automated High-Throughput Immunoassay Analyzer

ThyroidPublished Diagnostic accuracy study

Authors' conclusion

TSH from DBS may be accurately measured on an automated high-throughput immunoassay analyzer and could be used to diagnose hypothyroidism and, for the first time, hypothyrotropinemia.

Study context

Population
Patients presenting to the outpatient clinic of the University Medical Center Utrecht
Participants
109 patients
Comparison
Finger stick DBS TSH vs venous plasma TSH
Outcome
Method comparison of TSH: venous plasma vs finger stick DBS
Methodological detailsmethod comparison, single-centre, Validation of stability, hematocrit (Hct) effect, precision, and limits of blank and quantitation per Clinical and Laboratory Standards Institute evaluation protocol
  • Method comparison
  • Single-centre
  • Validation of stability, Hct effect, precision, and limits of blank and quantitation per Clinical and Laboratory Standards Institute evaluation protocol

Performance

As reported in the abstract

R2

0.988

95% CI 0.986–0.990

Correct diagnosis of hypothyrotropinemia

12 of 14

Correct diagnosis of hypothyroidism

6 of 7

Further findings

Stability and effect of Hct

TSH from DBS stable at −20°C to +30°C for at least
4 days
Maximum decrease of eluate TSH per 1% increase in Hct
1.13%

Original publication

A Preliminary Analysis of Thyrotropin Measurement from Finger Stick Dried Blood Spot with an Automated High-Throughput Immunoassay Analyzer. Thyroid. Published 15 March 2023. doi:10.1089/thy.2022.0453

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.