Diabetes technology

Testing the Real-World Accuracy of the Dexcom G6 Pro CGM During the Insulin-Only Bionic Pancreas Pivotal Trial

Diabetes Technol TherPublished Diagnostic accuracy study

Authors' conclusion

These results are comparable with accuracy shown previously with laboratory-based measurements and provide real-world evidence of Dexcom G6 Pro accuracy, which improved after the first 12 h and then remained stable thereafter.

Study context

Population
Participants with T1D in the Insulin-Only Bionic Pancreas Trial
Participants
53 participants1,073 CGM-BGM pairs
Comparison
Dexcom G6 Pro CGM vs blood glucose meter (BGM)
Outcome
Mean absolute relative difference (MARD) of CGM vs BGM
Follow-up
Median 50 h (range 47-79 h)
Methodological detailsblinded CGM, real-world setting, BGM comparator
  • Blinded CGM
  • Real-world setting
  • BGM comparator

Performance

As reported in the abstract

MARD of CGM vs BGM

Mean absolute relative difference (MARD) over a median period of 50 h

11.0%

Further findings

MARD by wear period

MARD, first 12 h
13.6%
MARD, hours 12-24
10.5%
MARD, after first 24 h
10.1%

Original publication

Testing the Real-World Accuracy of the Dexcom G6 Pro CGM During the Insulin-Only Bionic Pancreas Pivotal Trial. Diabetes Technol Ther. Published 3 October 2023. doi:10.1089/dia.2023.0287