Diabetes technology

Real-world Accuracy of Continuous Glucose Monitoring (CGM) in Inpatient Critical and Noncritical Care Settings at a Safety-Net Hospital

Diabetes CarePublished Retrospective cohort study

Authors' conclusion

Real-world accuracy of inpatient CGM is acceptable for NCI and CI patients. Further exploration of conditions associated with lower CGM accuracy in real-world settings is warranted.

Study context

Population
Hospitalized patients on Dexcom G6 CGM in a large safety-net hospital; 239 encounters (83 not critically ill (NCI), 156 CI)233 patients
Participants
2,744 vs 3,705
Exposure
CGM vs point of care (POC) and laboratory (Lab) glucose
  • POC-CGMn = 2,744
  • Lab-CGMn = 3,705
Outcome
Real-world accuracy of inpatient CGM (mean absolute relative differences (MARD), Clarke Error Grid zones A/B)
Methodological detailsreal-world, single safety-net hospital
  • Real-world
  • Single safety-net hospital

Main finding

As reported in the abstract

CGM accuracy

Aggregated MARD

Average MARD

Clarke Error Grid zones A/B

Observational data: associations, not effects.

Original publication

Real-world Accuracy of CGM in Inpatient Critical and Noncritical Care Settings at a Safety-Net Hospital. Diabetes Care. Published 10 August 2023. doi:10.2337/dc23-0089

Composed by AI from the published abstract and checked against it. How Briefs are made →